What does Biphasic Light Dose and Thermal Limits conclude, in numbers, about “Closing the Loop: a Device Parameter Sheet, the Light Dose Converter and a Pre-registered Test”?
Answer: Chapter 8 gives a device parameter sheet, the distilled Light Dose Converter and a pre-registered test plan for in vitro systems. Marked boxes separate published evidence, runnable Python calculations (18 programs), classical sources and the author's hypotheses.
Scope of the numbers: It writes only physical quantities (irradiance, fluence, temperature rise, duty cycle) and dose ranges reported in published studies; all devices and parameters are illustrative and all calculations are generated by its programs. It is not medical advice and claims no effect of any kind; anyone planning a self-experiment must first consult a licensed physician.
Data table
The numbers below are printed verbatim by the chapter's Python program and can be reproduced by running it. standard-library Python only
| setting | E mW/cm2 | F J/cm2 | Q J | T_end C | Omega | flags |
| probe 30 s | 353.7 | 10.61 | 3.0 | 37.0 | 7.22e-06 | - |
| probe 300 s | 353.7 | 106.12 | 30.0 | 57.6 | 9.42e+00 | HEAT OMEGA |
| cluster | 100.0 | 30.00 | 150.0 | 40.0 | 7.46e-05 | - |
| panel | 30.0 | 36.00 | 14400.0 | 36.0 | 2.43e-05 | - |
| pulsed 2 % | 100.0 | 30.00 | 30.0 | 40.0 | 7.46e-05 | - |
| weak pad | 5.0 | 3.00 | 3.0 | 33.5 | 3.55e-06 | - |
| eyes: lasers by IEC 60825-1 class, lamps and LEDs by IEC 62471 | ||||||
| risk group; goggles matched to the wavelength for any class 3B/4 |
Show the full program
# Chapter 8. p81: the device parameter sheet (no claims of any effect).
# For each setting: average irradiance E = duty * P_peak / area, fluence
# F = E * t / 1000, energy on the spot Q = duty * P_peak * t / 1000,
# then the lumped warming of p51 (T0 33 degC, a 0.5, h 0.005 W/cm^2/K,
# tau 252 s; average power, valid when the pulse period is far below
# tau, Ch. 6) and the Arrhenius integral of p41 over the session plus
# a 3 tau tail. Flags: HEAT if the end temperature exceeds 41 degC,
# OMEGA if the integral exceeds the illustrative margin 0.001.
# Every row is illustrative; the same formulas are in distill/tool.py.
import math
T0, ABS, H, C = 33.0, 0.5, 0.005, 1.26
TAU = C / H
A, EA, R = 3.1e98, 6.28e5, 8.314
MARGIN, T_FLAG = 0.001, 41.0
ROWS = [("probe 30 s", 100.0, 0.2827, 1.0, 30.0),
("probe 300 s", 100.0, 0.2827, 1.0, 300.0),
("cluster", 500.0, 5.0, 1.0, 300.0),
("panel", 12000.0, 400.0, 1.0, 1200.0),
("pulsed 2 %", 5000.0, 1.0, 0.02, 300.0),
("weak pad", 5.0, 1.0, 1.0, 600.0)]
def rate(tc):
return A * math.exp(-EA / (R * (tc + 273.15)))
def simpson(f, b, n=2000):
hs = b / n
terms = [f(0.0), f(b)]
for i in range(1, n):
terms.append((4.0 if i % 2 else 2.0) * f(i * hs))
return math.fsum(terms) * hs / 3.0
def sheet(p_peak, area, duty, t):
e = duty * p_peak / area
dss = ABS * e / 1000.0 / H
x_end = dss * (1.0 - math.exp(-t / TAU))
om = math.fsum([
simpson(lambda s: rate(T0 + dss * (1.0 - math.exp(-s / TAU))), t),
simpson(lambda s: rate(T0 + x_end * math.exp(-s / TAU)),
3.0 * TAU)])
flags = []
if T0 + x_end > T_FLAG:
flags.append("HEAT")
if om > MARGIN:
flags.append("OMEGA")
return e, e * t / 1000.0, duty * p_peak * t / 1000.0, \
T0 + x_end, om, " ".join(flags) or "-"
print("setting E mW/cm2 F J/cm2 Q J T_end C Omega flags")
for name, p, area, duty, t in ROWS:
e, f, q, te, om, fl = sheet(p, area, duty, t)
print("%-11s %8.1f %7.2f %6.1f %7.1f %.2e %s"
% (name, e, f, q, te, om, fl))
print("eyes: lasers by IEC 60825-1 class, lamps and LEDs by IEC 62471")
print("risk group; goggles matched to the wavelength for any class 3B/4")Single source and citation
Fan, Y. (2026). The Biphasic Dose Window of Photobiomodulation: Arrhenius Thermal Limits and Exposure Times under Pulsed Modulation (Version 1.0.0), Chapter 8: Closing the Loop: a Device Parameter Sheet, the Light Dose Converter and a Pre-registered Test. 5D Library. DOI pending
DOI pending · monograph page goes live after publication · 5DL-B-287 · v1.0.0 · CC-BY-NC-ND-4.0
BibTeX
@book{fan2026287,
author = {Fan, Yafu},
title = {The Biphasic Dose Window of Photobiomodulation: Arrhenius Thermal Limits and Exposure Times under Pulsed Modulation},
year = {2026},
publisher = {5D Library},
edition = {1.0.0},
note = {DOI pending}
}Other chapter answers in this book
- Chapter 1 · The Biphasic Dose Window: the Mathematical Form of the Arndt–Schulz Curve
- Chapter 2 · Irradiance Times Time: Converting Fluence, and the Failure of Reciprocity
- Chapter 3 · Optical Attenuation in Tissue: the Beer–Lambert Law and the Diffusion Approximation
- Chapter 4 · The Arrhenius Damage Integral: the Mathematics of a Temperature Boundary
- Chapter 5 · From Irradiance to Temperature Rise: the Thermal Boundary in the Irradiance and Time Plane
- Chapter 6 · Pulsed Modulation: Duty Cycle, Average Power and Temperature Swings
- Chapter 7 · East and West in the Sources: "Yang Qi Is Like the Sky and the Sun" and Schulz's Yeast