What does Biphasic Light Dose and Thermal Limits conclude, in numbers, about “Irradiance Times Time: Converting Fluence, and the Failure of Reciprocity”?
Answer: Chapter 2 converts among power, spot, irradiance and fluence and shows that reciprocity often fails: at the same 3 J/cm², different irradiances give different results, so dose needs two dimensions, irradiance and time.
Scope of the numbers: It writes only physical quantities (irradiance, fluence, temperature rise, duty cycle) and dose ranges reported in published studies; all devices and parameters are illustrative and all calculations are generated by its programs. It is not medical advice and claims no effect of any kind; anyone planning a self-experiment must first consult a licensed physician.
Data table
The numbers below are printed verbatim by the chapter's Python program and can be reproduced by running it. standard-library Python only
| device P mW area cm2 E mW/cm2 t to 3 J/cm2 Q J | |
| probe 0.6 cm | 100 0.2827 353.7 8.48 s 0.85 |
| probe 0.1 cm | 100 0.0079 12732.4 0.24 s 0.02 |
| cluster | 500 5.0000 100.0 30.00 s 15.00 |
| panel | 12000 400.0000 30.0 100.00 s 1200.00 |
| weak pad | 5 1.0000 5.0 600.00 s 3.00 |
| the same 4 J delivered on each spot, read as a fluence: | |
| probe 0.6 cm | 4 J / 0.2827 cm2 = 14.15 J/cm2 |
| probe 0.1 cm | 4 J / 0.0079 cm2 = 509.30 J/cm2 |
| cluster | 4 J / 5.0000 cm2 = 0.80 J/cm2 |
| panel | 4 J / 400.0000 cm2 = 0.01 J/cm2 |
| weak pad | 4 J / 1.0000 cm2 = 4.00 J/cm2 |
| shrinking the spot from 0.6 to 0.1 cm multiplies E by 36 |
Show the full program
# Chapter 2. p21: power, spot area, irradiance, fluence and energy.
# Definitions: irradiance E = P / area (mW/cm^2); fluence (energy
# density) F = E * t / 1000 (J/cm^2, t in s); energy on the spot
# Q = P * t / 1000 (J). A round spot of diameter d cm has area
# pi * d^2 / 4. The four devices are illustrative, not products.
# Second table: the same "4 J per point" read as a fluence depends on
# the spot area, which is why J and J/cm^2 must never be swapped.
import math
DEVICES = [
("probe 0.6 cm", 100.0, math.pi * 0.6 ** 2 / 4.0),
("probe 0.1 cm", 100.0, math.pi * 0.1 ** 2 / 4.0),
("cluster", 500.0, 5.0),
("panel", 12000.0, 400.0),
("weak pad", 5.0, 1.0),
]
TARGET = 3.0 # J/cm^2, an illustrative fluence for the arithmetic only
print("device P mW area cm2 E mW/cm2 t to 3 J/cm2 Q J")
for name, p, area in DEVICES:
e = p / area
t = TARGET * 1000.0 / e
q = p * t / 1000.0
print("%-13s %7.0f %9.4f %9.1f %11.2f s %6.2f"
% (name, p, area, e, t, q))
print("the same 4 J delivered on each spot, read as a fluence:")
for name, p, area in DEVICES:
print(" %-13s 4 J / %8.4f cm2 = %8.2f J/cm2"
% (name, area, 4.0 / area))
d0, d1 = 0.6, 0.1
print("shrinking the spot from %.1f to %.1f cm multiplies E by %.0f"
% (d0, d1, (d0 / d1) ** 2))Single source and citation
Fan, Y. (2026). The Biphasic Dose Window of Photobiomodulation: Arrhenius Thermal Limits and Exposure Times under Pulsed Modulation (Version 1.0.0), Chapter 2: Irradiance Times Time: Converting Fluence, and the Failure of Reciprocity. 5D Library. DOI pending
DOI pending · monograph page goes live after publication · 5DL-B-287 · v1.0.0 · CC-BY-NC-ND-4.0
BibTeX
@book{fan2026287,
author = {Fan, Yafu},
title = {The Biphasic Dose Window of Photobiomodulation: Arrhenius Thermal Limits and Exposure Times under Pulsed Modulation},
year = {2026},
publisher = {5D Library},
edition = {1.0.0},
note = {DOI pending}
}Other chapter answers in this book
- Chapter 1 · The Biphasic Dose Window: the Mathematical Form of the Arndt–Schulz Curve
- Chapter 3 · Optical Attenuation in Tissue: the Beer–Lambert Law and the Diffusion Approximation
- Chapter 4 · The Arrhenius Damage Integral: the Mathematics of a Temperature Boundary
- Chapter 5 · From Irradiance to Temperature Rise: the Thermal Boundary in the Irradiance and Time Plane
- Chapter 6 · Pulsed Modulation: Duty Cycle, Average Power and Temperature Swings
- Chapter 7 · East and West in the Sources: "Yang Qi Is Like the Sky and the Sun" and Schulz's Yeast
- Chapter 8 · Closing the Loop: a Device Parameter Sheet, the Light Dose Converter and a Pre-registered Test